Measurement of fetal heart rate variability on an electronic monitor using a prototype electronic ruler
Lianna M Hall1, Daniel J Hannon, Gary G Leisk
1Department of Mechanical Engineering, Tufts University, Medford, MA, USA. liannamhall@gmail.com
Objective:
To develop a prototype electronic ruler for assessment of fetal heart rate (FHR) variability on an electronic monitor and test its reliability and accuracy.
Study Design:
A prototype electronic ruler was designed and developed for assessment of FHR variability on electronic monitors. The electronic ruler consisted of horizontal bands that were sized and colored to embed the four FHR variability categories. The reliability and accuracy of using the electronic ruler to assess FHR variability was studied with expert clinicians.
Results:
Intrarater and interrater reliability was moderate for both the electronic ruler and paper strips. The amplitude measurement accuracy of expert variability assessment compared with a gold standard was significantly improved (p < 0.001) with the electronic ruler versus paper strips. The accuracy of subjects' FHR variability category responses compared with the gold standard revealed no significant difference (p = 0.50) using either display type.
Conclusion:
Performance of the electronic ruler was equivalent to paper strips, which may aid assessment of variability on electronic monitors as paper strips become less prevalent.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Dysrhythmias V: Evaluating Dysrhythmias
Holter Monitor: 24-Hour Monitoring
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage. When...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: