Related Experiment Video
Updated: May 24, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Development and validation of an open screening method for diuretics, stimulants and selected compounds in human
A Jiménez Girón1, K Deventer, K Roels
1Department of Analytical Chemistry, University of Extremadura, Badajoz, Spain. ajimenezgiron@gmail.com
Abstract:
A new doping control screening method for the analysis of diuretics and stimulants using ultra high pressure liquid chromatography-high resolution Orbitrap mass spectrometry has been developed. The screening was performed in full scan MS with scan-to-scan polarity switching which allowed to detect more than 120 target analytes. Sample preparation was limited to 10-fold dilution of the urine into the internal standard solution followed by injection. Total run time per sample was 10 min. Validation of the method yielded detection limits for diuretics between 25 and 250 ng mL(-1) and for stimulants between 5 and 500 ng mL(-1). The screening method has been implemented in routine doping control.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry (UPLC-HRMS)
Published on: May 20, 2013
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Urine Studies I: Urinalysis