Biomechanical analysis of locked and non-locked plate fixation of the clavicle
K J Little1, P E Riches, U G Fazzi
1Department of Orthopaedic Surgery, Western Infirmary, Dumbarton Road, Glasgow G11 6NT, United Kingdom. kevinlittle1977@gmail.com
Introduction:
During plate fixation of clavicular fractures the brachial plexus and subclavian vessels are vulnerable to injury beneath the clavicle. Locking plate fixation allows for mono-cortical fixation, theoretically reducing the risk of injury to these structures. Biomechanical analysis of the performance of such fixation is limited, and this study was designed to explore this further as a treatment option in clavicle fractures.
Materials And Methods:
Fixation of fifteen simulated mid-shaft fractures was undertaken using a combination of mono-cortical locked, bicortical locked and bicortical non-locked plating methods in cadaveric clavicles. Samples were then tested via three-point bending to destruction, and the performance of each with respect to failure load, bending stress, bending stiffness and Young's modulus was then analysed. The influence of the number of cortices engaged and locking was also assessed.
Results:
Clavicles fixed with monocortical locking plates displayed a significantly lower bending stress (12±1 MPa) than both the bicortical locking (28±3 MPa, p=0.015) and non-locking specimens (24±3 MPa, p=0.002). Engaging two cortices with the fixation produced a significant increase in failure load (291±28 N vs 138±48 N, p=0.018) and bending stress (26±2 MPa vs 9.9±3.5 MPa, p=0.002) compared to single cortex fixation.
Discussion:
The greatest influence upon the performance of the fixation was the number of cortices engaged, with bicortical fixation performing significantly better than mono-cortical. Whether or not the fixation device was a locking one did not have a significant bearing upon the performance.
Conclusion:
This in vitro biomechanical analysis demonstrates that mono-cortical locked plating fails at significantly lower levels of load and stress than bicortical locked and non-locked plating in mid-shaft fractures of the clavicle, and caution would therefore be advised in its use as a fixation modality for these injuries.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Related Concept Videos
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...
Changes in the Appendicular Skeleton with Age
Initially, the limb buds consist of a core of mesenchyme covered by a layer of ectoderm. The ectoderm at the end of the limb bud thickens to form a narrow crest called the apical ectodermal ridge. This ridge stimulates the underlying...