Multi-dimensional gas chromatography with a planar microfluidic device for the characterization of volatile
1Australian Centre for Research on Separation Science (ACROSS), University of Tasmania, Private Bag 75, Hobart, Tasmania 7001, Australia.
Abstract:
Oxygenated compounds like methanol, ethanol, 1-propanol, 2-propanol, 1-butanol, acetaldehyde, crotonaldehyde, ethylene oxide, tetrahydrofuran, 1,4-dioxane, 1,3-dioxolane, and 2-chloromethyl-1,3-dioxolane are commonly encountered in industrial manufacturing processes. Despite the availability of a variety of column stationary phases for chromatographic separation, it is difficult to separate these solutes from their respective matrices using single dimension gas chromatography. Implemented with a planar microfluidic device, conventional two-dimensional gas chromatography and the employment of chromatographic columns using dissimilar separation mechanisms like that of a selective wall-coated open tubular column and an ionic sorbent column have been successfully applied to resolve twelve industrially significant volatile oxygenated compounds in both gas and aqueous matrices. A Large Volume Gas Injection System (LVGIS) was also employed for sample introduction to enhance system automation and precision. By successfully integrating these concepts, in addition to having the capability to separate all twelve components in one single analysis, features associated with multi-dimensional gas chromatography like dual retention time capability, and the ability to quarantine undesired chromatographic contaminants or matrix components in the first dimension column to enhance overall system cleanliness were realized. With this technique, a complete separation for all the compounds mentioned can be carried out in less than 15 min. The compounds cited can be analyzed over a range of 250 ppm (v/v) to 100 ppm (v/v) with a relative standard deviation of less than 5% (n=20) with high degree of reliability.
More Related Videos
10:14Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...