Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Comparison of cone-beam and conventional multislice computed tomography for image-guided dental implant planning
Paul W Poeschl1, Nina Schmidt, Godoberto Guevara-Rojas
1University Hospital for Cranio-Maxillofacial and Oral Surgery, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria.
Objectives:
To compare the accuracy of cone-beam CT (CBCT) and multislice CT (MSCT) with regard to its use in image-guided dental implant surgery in a prospective model based study.
Material And Methods:
Ten photopolymer-acrylate mandibula models, each with four precise metal reference markers, were scanned with MSCT and CBCT. The six reference distances between the markers were measured by a three-axis milling machine first. The distances were then measured by (1) navigation with the Medtronic StealthStation® TREON™ image-guided surgery system, (2) with the Medtronic planning-tool and (3) on the PC with the Mimics® software. Mean values were calculated for all three methods for CBCT and MSCT and were compared for statistical significance.
Results:
Of all measurements, 83% of the arithmetic mean values were within the ±0.5 mm range (MSCT 88% and CBCT 78%) and 17% within the ±1.0 mm range (MSCT 12% and CBCT 22%). The absolute difference of the arithmetic mean values showed no statistically significant difference between MSCT and CBCT. The difference of the overall mean values to the reference was 0.43 mm for MSCT and 0.46 mm for CBCT.
Conclusions:
The data of our study prove that the application of CBCT for the indicated purpose yielded good results comparable to those of MSCT. All three measuring methods were feasible and accuracy was statistically not different between the data acquired by MSCT and CBCT within the setting of this study.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...