Agreement between direct, oscillometric and Doppler ultrasound blood pressures using three different cuff positions
Natache A Garofalo1, Francisco J Teixeira Neto, Renata K Alvaides
1Departamento de Cirurgia e Anestesiologia Veterinária, Faculdade de Medicina Veterinária e Zootecnia, Universidade Estadual Paulista, São Paulo, Brazil.
Objective:
To evaluate the agreement between invasive blood pressure (IBP) and Doppler ultrasound blood pressure (DUBP) using three cuff positions and oscillometric blood pressure (OBP) in anesthetized dogs.
Study Design:
Prospective study.
Animals:
Nine adult dogs weighing 14.5-29.5 kg.
Methods:
The cuff was placed above and below the tarsus, and above the carpus with the DUBP and above the carpus with the OBP monitor. Based on IBP recorded via a dorsal pedal artery catheter, conditions of low, normal, and high systolic arterial pressures [SAP (mmHg) <90, between 90 and 140, and >140, respectively] were induced by changes in isoflurane concentrations and/or dopamine administration. Mean biases ± 2 SD (limits of agreement) were determined.
Results:
At high blood pressures, regardless of cuff position, SAP determinations with the DUBP underestimated invasive SAP values by more than 20 mmHg in most instances. With the DUBP, cuff placement above the tarsus yielded better agreement with invasive SAP during low blood pressures (0.2 ± 16 mmHg). The OBP underestimated SAP during high blood pressures (-42 ± 42 mmHg) and yielded better agreement with IBP for mean (MAP) and diastolic (DAP) arterial pressure measurements [overall bias: 2 ± 15 mmHg (MAP) and 0.2 ± 16 mmHg (DAP)].
Conclusions:
Agreement of SAP determinations with the DUBP is poor at SAP > 140 mmHg, regardless of cuff placement. Measurement error of the DUBP with the cuff placed above the tarsus is clinically acceptable during low blood pressures. Agreement of MAP and DAP measurements with this OBP monitor compared with IBP was clinically acceptable over a wide pressure range.
Clinical Relevance:
With the DUBP device, placing the cuff above the tarsus allows reasonable agreement with IBP obtained via dorsal pedal artery catheterization. Only MAP and DAP provide reasonable estimates of direct blood pressure with the OBP monitor evaluated.
More Related Videos
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: