Related Experiment Video
Updated: May 24, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Relation between projection positions and X-ray radiation doses during placement of the coronary sinus mapping
Gang Lin1, Jian-Fei Huang, Hui-He Lu
1Cardiovascular Medical Department, The First People's Hospital of Nantong, Nantong 226001, China.
Purpose:
To reduce X-ray radiation injury, we investigated the relation between projection positions and X-ray radiation doses when a coronary sinus mapping electrode was placed.
Methods:
There were 400 patients in whom a coronary sinus mapping electrode was placed for intracardiac electrophysiologic examination or radio-frequency catheter ablation in our hospital between 1 June 2008 and 31 May 2011. The doctors who placed the coronary sinus mapping electrode and the projection positions including posteroanterior, left anterior oblique 45°, and right anterior oblique 30° projection positions were randomly selected. X-ray doses were recorded when the coronary sinus mapping electrode was placed, and then, these were compared between the three projection positions.
Results:
The X-ray dose was significantly lower in the right anterior oblique 30° projection position (50.01 ± 11.38 mGy) than in the posteroanterior projection position (68.05 ± 18.85 mGy, q = 12.924, p = 0.000) and the left anterior oblique 45° projection position (71.66 ± 16.95 mGy, q = 15.214, p = 0.000). There were no statistical significances in X-ray doses between the posteroanterior and the left anterior oblique 45° projection positions (p = 0.066) and different operators (p = 0.985).
Conclusion:
Based on the results of this study, we suggest that in placing the coronary sinus mapping electrode, the right anterior oblique 30° projection position should be first adopted because it can allow the electrode to be easily placed and reduce the X-ray radiation dose as much as possible.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...