Measurement of anisotropic angular distributions of photon energy spectra for I-125 brachytherapy sources
Yasuhiro Unno1, Akira Yunoki, Tadahiro Kurosawa
1National Metrology Institute of Japan, Tsukuba, Ibaraki, Japan. y.unno@aist.go.jp
Abstract:
The angular distribution of photon energy spectra emitted from an I-125 brachytherapy source was measured using a specially designed jig in the range of ±70° in the plane of the long axis of the source. It is important to investigate the angular dependence of photon emissions from these sources for the calibration of the air kerma rate. The results show that the influence of the distributions between 0° and ±8° is small enough to allow a calibration using current primary instruments which have a large entrance window.