Single-incision apical and posterior mesh repair: 1-year prospective outcomes
James C Lukban1, Jan-Paul W R Roovers, Douglas M Vandrie
1Division of Urogynecology, Department of Obstetrics and Gynecology, Eastern Virginia Medical School, 825 Fairfax Avenue, Suite 310, Norfolk, VA 23507, USA. lukbanjc@evms.edu
Introduction And Hypothesis:
The objective of this study was to assess the safety and efficacy of the Elevate Apical and Posterior single-incision mesh system (SIMS) with IntePro Lite for pelvic organ prolapse repair.
Methods:
This prospective multicenter study included 139 women with ≥ stage II posterior vaginal prolapse and/or apical descent who underwent placement of type I polypropylene mesh through a single transvaginal incision with no external needle passes. Primary endpoint was the percent of patients with posterior and/or apical stage ≤ I ("cure") at follow-up. Secondary endpoints included, but were not limited to, rate of mesh extrusion and disease-specific quality of life outcomes.
Results:
At 12 months, objective posterior wall and apical cure rates were 92.5 and 89.2 %, respectively, with an extrusion rate of 6.5 %.
Conclusions:
The SIMS appears to be effective and safe in treating patients with posterior vaginal and/or apical prolapse. The risks and benefits of transvaginal synthetic mesh insertion should be considered.