Calibration of stack monitors for measurement of noble gases in nuclear facilities
Petr Kovar1, Pavel Dryak, Jiri Suran
1Czech Metrology Institute, Inspectorate for Ionizing Radiation, Radiova 1, CZ-102 00 Prague 10, Czech Republic. pkovar@cmi.cz
Abstract:
In nuclear facilities stack monitors are used for the measurement of the volumetric activity of noble gases. Spectrometric measurement is needed because the content of stack effluents is always a mixture of radionuclides. In some nuclear power plants new types of monitors were installed based on HPGe detectors. For efficiency calibration a standard with the radionuclide Xe-127 was developed and calibration curve constructed in the energy range 81 keV-1293 keV. Experiental efficiencies were checked using an MC model.
More Related Videos
Related Concept Videos
Noble Gases
The elements in group 18 are noble gases (helium, neon, argon, krypton, xenon, and radon). They earned the name “noble” because they were assumed to be nonreactive since they have filled valence shells. In 1962, Dr. Neil Bartlett at the University of British Columbia proved this assumption to be false.