Related Experiment Video
Updated: May 24, 2026

An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
In vitro activation of human leukocytes in response to contact with synthetic hernia meshes
1Clinical Engineering, UKCTE, UKBioTEC, The Institute of Ageing and Chronic Disease, University of Liverpool, Duncan Building, Daulby Street, Liverpool, L69 3GA, UK. n.bryan@liv.ac.uk
Objectives:
Evaluation of an in vitro chemiluminescent screen to predict leukocyte ROS in response to surgical materials.
Design And Methods:
6 surgical meshes; manufacture and knitting variations of polypropylene (PP), polyester terephtalate (PET) and polyglycolic acid (PGA) trialled healthy human blood (n=5). Materials and blood were incubated with pholasin. Pholasin emits photons in the presence of reactive oxygen species; secreted by activated leukocytes.
Results:
Multifilament-PGA mesh stimulated the greatest ROS response from blood derived human leukocytes. Multifilament-PET light weight and multifilament-PP meshes stimulated similar levels of ROS production which were greater than monofilament-PP light, monofilament-PP and monofilament-PET light meshes. Data demonstrated statistical variations in trans-donor response to the materials.
Conclusions:
An in vitro chemiluminescent assay can be used to assess leukocyte respiratory burst response to biomaterials. PGA mesh elicited the greatest ROS response. PP and PET monofilament meshes induce less ROS than multifilament equivalents. In vitro results correlate with previously published clinical responses to these materials.