Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
Automatic office blood pressure measured without doctors or nurses present
Joji Ishikawa1, Efthimia G Nasothimiou, Nikos Karpettas
1Center for Behavioral Cardiovascular Health, Division of General Medicine, Columbia University Medical Center, New York, New York, USA. george@jichi.ac.jp
Background:
We evaluated the agreement between office blood pressure (OBP) measured by a mercury sphygmomanometer (Sphyg) and an automatic (Auto) device without any observers, and compared Auto and Sphyg OBP with ambulatory blood pressure (ABP) and home blood pressure (HBP).
Methods:
OBP was measured in 75 hypertensive patients at two sites using an automatic monitor without a doctor or a nurse present and by Sphyg during three clinic visits. Between visits, monitoring of ABP and HBP was also performed.
Results:
The mean Auto OBP was similar to that of Sphyg OBP and the values were closely correlated (intraclass correlation coefficient=0.84 for systolic OBP and 0.91 for diastolic OBP); however, the difference between Auto and Sphyg systolic OBP (1.6 ± 8.2 mmHg) varied by the first office visit, sex, and the site. Auto systolic OBP was lower than both systolic awake ABP (137.1 ± 14.7 mmHg) and HBP (139.2 ± 15.6 mmHg). Auto systolic OBP and Sphyg OBP were similarly correlated with systolic awake ABP (both r=0.59, P<0.001). The mean Auto diastolic OBP was similar to that of Sphyg OBP (81.1 ± 11.3 vs. 80.3 ± 13.3 mmHg, P=0.20, intraclass correlation coefficient=0.91), and diastolic awake ABP and HBP. Auto diastolic OBP and Sphyg OBP were related to diastolic awake ABP (both r>0.68, P<0.001). In multivariable analyses, neither OBP measure was a significantly stronger predictor of out-of-office blood pressure than the other.
Conclusion:
Auto systolic OBP measured without a doctor or a nurse present was lower than systolic awake ABP and HBP. Auto and rigorously assessed Sphyg OBP had similar means and were similarly related to awake ABP. Auto OBP might be an advantageous alternative to Sphyg measurements in the usual clinic setting.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.