Effect of recording site on pulse oximetry readings
Emma Barnett1, Annette Duck, Richard Barraclough
1Wythenshawe Hospital.
Background:
Ambulatory oxygen requirements are routinely assessed and titrated using portable finger pulse oximetry. However, movement artefacts from hands moving or reduced circulation may adversely affect the accuracy of measurement. At Wythenshawe Hospital, we decided to use an ear oximeter routinely in addition to the finger oximeter.
Aim:
To evaluate and compare the use of portable oximetry measurements taken from the ear and finger during ambulatory assessments.
Method:
Before, during and after a six-minute walking test, 304 patients had ear and finger pulse oximetry measurements recorded.
Results:
There was a significant difference between ear and finger measurements before and after exercise (p=<0.001 for both). The differences would have altered the clinical outcome in 27% of those assessed.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Uncertainty in Measurement: Reading Instruments
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Regulation of Pulse