Related Experiment Video
Updated: May 23, 2026

06:04
Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Transcatheter aortic bioprosthesis dislocation: technical aspects and midterm follow-up
Gian Paolo Ussia1, Marco Barbanti, Kunal Sarkar
1Division of Cardiology, Ferrarotto Hospital, University of Catania, Catania, Italy. gpussia@hotmail.com
Summary
Prosthesis dislocation during transcatheter aortic valve implantation (TAVI) is rare. Dislocated CoreValve ReValving™ System (CRS) devices functioned normally, and a second valve implantation was safe and effective.
Area of Science:
- Cardiovascular Surgery
- Interventional Cardiology
- Medical Device Technology
Background:
- Prosthesis dislocation is a rare but significant complication during transcatheter aortic valve implantation (TAVI).
- Limited data exists on the performance of dislocated aortic valve prostheses.
- Investigating dislocation mechanisms of the CoreValve ReValving™ System (CRS) is crucial.
Purpose of the Study:
- To investigate the causes of CoreValve ReValving™ System (CRS) prosthesis dislocation during TAVI.
- To assess the midterm outcomes of patients experiencing prosthesis dislocation.
- To analyze the performance of dislodged CRS devices in the aorta.
Main Methods:
- Retrospective analysis of 176 consecutive TAVI procedures using the CRS prosthesis.
- Classification of dislocation mechanisms into accidental, repositioning-related, and intentional bailout.
- Midterm clinical follow-up of patients with dislocated CRS devices.
Main Results:
- Seven patients (3.9%) experienced acute CRS dislocation.
- Dislocations occurred during early use of the Accutrak™ delivery system.
- Dislodged devices functioned normally without migration or structural deterioration; no thromboembolic events occurred.
Conclusions:
- TAVI prosthesis dislocation can be managed by implanting a second device, leaving the dislodged device safely in the aorta.
- Intentional repositioning of dislodged CRS devices is feasible and safe for experienced operators.
- Midterm follow-up shows no serious adverse events related to managed prosthesis dislocations.