A Bedside, Single Burr Hole Approach to Multimodality Monitoring in Severe Brain Injury
Published on: March 26, 2019
Anchoring depth electrodes for bedside removal: a "break-away" suturing technique for intracranial monitoring
Robert E Gross1, Nathan C Rowland, Edward K Sung
1Department of Neurosurgery, Emory University School of Medicine, Atlanta, Georgia 30322, USA. rgross@emory.edu
Background:
Intracranial depth electrodes for epilepsy are easily dislodged during long-term monitoring unless adequately anchored, but a technique is not available that is both secure and allows easy explantation without reopening the incision.
Objective:
To describe a convenient and inexpensive method for anchoring depth electrodes that prevents migration and incidental pullout while allowing electrode removal at the bedside.
Methods:
An easily breakable suture (eg, MONOCRYL) is tied around both the depth electrode and a heavy nylon suture and anchored to a hole at the edge of the burr hole; the tails of both are tunneled together percutaneously. The "break-away" MONOCRYL suture effectively anchors the electrode for as long as needed. At the completion of the intracranial electroencephalography session, the 2 tails of the nylon suture are pulled to break their encompassing MONOCRYL anchor suture, thus freeing the depth electrode for easy removal.
Results:
The break-away depth electrode anchoring technique was used for 438 electrodes in 68 patients, followed by explantation of these and associated strip electrodes without reopening the incision. Only 1 electrode (0.2%) migrated spontaneously, and 3 depth electrodes (0.7%) fractured in 2 patients (2.9%) on explantation, necessitating open surgery to remove them in 1 of the patients (1.5%).
Conclusion:
An easy and inexpensive anchoring configuration for depth electrodes is described that provides an effective and safe means of securing the electrodes while allowing easy explantation at the bedside.