Pullout analysis of a lumbar plate with varying screw orientations: experimental and computational analyses
David Rios1, Wyzscx Merfil Patacxil, Daniel Kyle Palmer
1Department of Orthopaedic Surgery, Orthopaedic Biomechanics Laboratory, Loma Linda University, Loma Linda, CA 92354, USA.
Study Design:
Experimental and finite element analysis of anterior lumbar interbody fixation (ALIF) plate pullout.
Objective:
The objective of this study was to determine the effect of screw angle and orientation on ALIF plate pullout strength.
Summary Of Background Data:
It has been thought that angling the screws in an ALIF plate leads to better fixation strength; however, a few studies are published on this question, which produced conflicting results.
Methods:
Using custom guides, screws were configured in 9 different orientations to affix ALIF plates to polyurethane foam blocks. Pullout tests were performed at a rate of 1 mm/min. In addition, finite element analyses were performed on a 2-dimensional screw-block model to gain insight into the internal stress during pullout.
Results:
The pullout load was the greatest, with screws positioned 12° outward sagittaly and 6° inward coronally (936 ± 72 N). This orientation was statistically greater than the orientation with the lowest pullout load (812 ± 45 N, P < 0.05); however, no group was statistically different than placing the screws straight in (868 ± 86 N, P > 0.05). Finite elements analysis showed some gain in pullout strength at 12° followed by some loss at greater angles. As the screw insertion angle increased, stress levels elevated within the block even in the regions away from the screw.
Conclusion:
Significant difference was found between certain screw-angle configurations; however, when compared with simply placing the screws straight in, the difference was never more than 8%. This implies that there is greater freedom in the angle and placement of screws than previously thought. Our results show that there is little change in fixation strength when placing the screw in a different direction.
Related Concept Videos
Plastic Deformations of Members with a Single Plane of Symmetry
Eccentric Axial Loading in a Plane of Symmetry
Screw: Problem Solving
To evaluate the...
Plastic Deformations
Residual Stresses in Circular Shafts
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical bending,...