Bacterial contamination and cleanliness of emergency department ultrasound probes
Geoffrey E Sanz1, Jonathan Theoret, Michael M Liao
1Department of Emergency Medicine, Queen’s University, Kingston, ON.
Objectives:
As ultrasonography is increasingly used in the emergency department (ED), ultrasound equipment has become a potential threat to infection control. Improperly cleaned ultrasound probes may serve as a vector for pathogens such as methicillin-resistant Staphylococcus aureus (MRSA). The primary objective of this study was to determine the prevalence of MRSA colonization on ultrasound probes used in a busy, urban ED. It was hypothesized that cultures of our ED ultrasound probes would yield a significant number of positive results for MRSA.
Methods:
In this observational study, 11 ED ultrasound probes were randomly sampled on 10 different occasions. Samples were taken using a RODAC plate method and were cultured for MRSA and methicillin-sensitive Staphylococcus aureus (MSSA). On half of the randomly assigned sampling occasions, a visual inspection of each ultrasound probe for general cleanliness was conducted and recorded. Data were stratified by ultrasound location in the ED and analyzed using the Fisher exact test, with p < 0.05 deemed to be statistically significant.
Results:
Of 110 samples, no isolates of MRSA were cultured. One probe yielded a positive culture for MSSA. Probes in the medicine, trauma, and pediatrics areas were found to be clean 65%, 33%, and 70% of the time, respectively. This variability in probe cleanliness by ED location was found to be statistically significant (p < 0.01).
Conclusions:
Contrary to our hypothesis, MRSA contamination of ultrasound probes was not found. This finding suggests that the spread of MRSA by ED ultrasound machines in a high-volume urban ED is unlikely. Further research at different centres with larger sample sizes is required before these results can be generalized.
Insights
This study found no methicillin-resistant Staphylococcus aureus (MRSA) on emergency department ultrasound probes, suggesting MRSA spread via these devices is unlikely in this setting. Further research is needed for broader conclusions.
Area of Science:
- Infection Control
- Medical Microbiology
- Diagnostic Imaging
Background:
- Ultrasonography is common in emergency departments (EDs).
- Ultrasound probes can pose infection risks if not properly cleaned.
- Pathogens like methicillin-resistant Staphylococcus aureus (MRSA) could potentially spread via contaminated probes.
Purpose of the Study:
- To determine the prevalence of MRSA colonization on ultrasound probes in a busy, urban ED.
- To assess the risk of MRSA transmission through ED ultrasound equipment.
Main Methods:
- An observational study involving 11 ED ultrasound probes sampled 10 times each.
- Samples cultured for MRSA and methicillin-sensitive Staphylococcus aureus (MSSA) using RODAC plates.
- Visual cleanliness inspections recorded; data stratified by ED location and analyzed with Fisher exact test.
Main Results:
- No MRSA isolates were cultured from 110 samples.
- One probe tested positive for MSSA.
- Probe cleanliness varied significantly by ED location (medicine 65%, trauma 33%, pediatrics 70%).
Conclusions:
- Contrary to hypothesis, no MRSA contamination was detected on ED ultrasound probes.
- MRSA spread via ED ultrasound machines in this high-volume urban ED appears unlikely.
- Further multi-center research with larger sample sizes is recommended for generalizability.
More Related Videos
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Handwashing III: During the Procedure and Post-Procedure Steps
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...