Programmable valve shunts: are they really better?
Rashim Kataria1, Vimal Kumar, Veer Singh Mehta
1Paras Hospital, Department of Neurosurgery, Gurgaon, Haryana, India. rashim_kat@yahoo.com
Abstract:
Programmable valve shunts allows selection of opening pressure of shunt valve. In the presented article, a unique complication pertaining to programmable shunts has been discussed. A 5-year-old boy who had tectal plate low grade glioma with obstructive hydrocephalus was managed with Codman programmable ventriculoperitoneal shunt. There was a spontaneous change in the opening pressure of the shunt valve leading to shunt malfunction. Routinely used household appliances produce a magnetic field strong enough to cause change in the setting of shunt valve pressure and may lead to valve malfunction. Other causes of programmable valve malfunction also discussed.
Related Concept Videos
Control of Power Flow
Underflow Gates
Shunt Admittances
To model this effect, the method of images is employed. This method involves replacing the Earth with an image conductor that mirrors the original...
Multiple Pipe Systems
Series Configuration
In a series configuration, fluid flows sequentially from one pipe...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...