Related Experiment Video
Updated: May 23, 2026

Augmented Reality Navigation-Guided Core Decompression for Osteonecrosis of Femoral Head
Published on: April 12, 2022
Accelerometer-based computer navigation for performing the distal femoral resection in total knee arthroplasty
Denis Nam1, Danyal H Nawabi, Michael B Cross
1Adult Reconstruction and Joint Replacement Service, Department of Orthopaedic Surgery, Hospital for Special Surgery, Weill Medical College of Cornell University, New York, New York, USA.
Abstract:
The use of intramedullary alignment guides for performing the distal femoral resection in total knee arthroplasty (TKA) have not been shown to be highly accurate. Forty-eight knees underwent a TKA using a portable, accelerometer-based surgical navigation system for performing the distal femoral resection (KneeAlign 2 system; OrthAlign, Inc, Aliso Viejo, Calif). Of the femoral components, 95.8% were placed within 90° ± 2° to the femoral mechanical axis in the coronal plane, and 93.8% of the TKAs had an overall lower extremity alignment within 3° of neutral to the mechanical axis, based on postoperative, standing, hip-to-ankle radiographs. The KneeAlign 2 is highly accurate in positioning the femoral component in TKA, and accelerometer-based navigation is able to reliably determine the hip center of rotation and femoral mechanical axis.