Related Experiment Video
Updated: May 23, 2026

Robotic Ablation of Atrial Fibrillation
Published on: May 29, 2015
Robotically controlled ablation for atrial fibrillation: the first real-world experience in Africa with the Hansen
Faizel Lorgat1, Evan Pudney, Helena van Deventer
1Department of Cardiology, Christiaan Barnard Memorial Hospital, Cape Town, South Africa. florgat@iafrica.com
Background:
We report the first single-centre experience in Africa with the Sensei X robotic navigation system in an unselected subset of patients with atrial fibrillation (AF).
Methods:
Data were recorded prospectively of all consecutive patients who underwent robotically assisted catheter ablation therapy using the Sensei X robotic navigation system at the Christiaan Barnard Memorial Hospital, Cape Town, South Africa, from July 2009 to July 2010. Outcomes were defined at one and nine months.
Results:
A total of 95 patients were included: 63% had only AF and 37% had AF plus atrial flutter. AF was of the persistent type in 81% of patients. The mean procedure, fluoroscopy and ablation times were 220.6 ± 89.6 min, 31.0 ± 20.4 min, and 61.3 ± 28.1 min, respectively. Both fluoroscopy and procedure times were significantly longer for the first 19 patients compared with the remaining 76 patients (43.5 ± 22.7 vs 27.8 ± 18.5 min and 274.7 ± 90.2 vs 207.1 ± 84.7 min, respectively, p = 0.002). The procedural endpoint of the study was successfully achieved in all patients. After one attempt, 27% were discharged from hospital off anti-arrhythmic drugs (AADs). At a median of nine months' follow up, 74% were AF-free off AADs, and 11% were AF-free on AADs, yielding a total freedom from AF of 84% without any redo procedures. Freedom from relapse after 1.12 procedures was 88%.
Conclusion:
The Sensei X robotic navigation system offers a safe and effective approach for the treatment of AF. There was a learning curve with regard to fluoroscopy and procedure time, after which point reduction in radiation exposure and operator strain, as well as improvement in procedure throughputs were even more pronounced.