A 5-kg time-resolved luminescence photometer with multiple excitation sources
1U. S. Department of Agriculture, Agricultural Research Service, Eastern Regional Research Center, 600 E. Mermaid Lane, Wyndmoor, Pennsylvania 19038, USA. guoying.chen@ars.usda.gov
Abstract:
A luminescence photometer was developed based on lanthanide-sensitized luminescence to detect environmental pollutants and residues in foods including, in particular, two classes of antibiotics: tetracyclines and fluoroquinolones. Multiple excitation sources, a xenon flashlamp and ultraviolet light-emitting diodes (UV LEDs), were used to their advantages. A photomultiplier tube module, gated to reject time-domain interferences, was used as a photodetector. Using danofloxacin as a model analyte, luminescence signal was linear in more than three decades (0.5-2000 ng/mL) with R(2) > 0.998 in each decade. This photometer achieved a limit of detection of 2.0 ng/mL with 1.3% average relative standard deviation. It is field deployable at 4.6 kg and 15 W power consumption.