Alphafetoprotein in the Dutch External Quality Assurance programme: a need for improvement
A Carolien Houwert1, M T W Tycho Lock, Eef G W M Lentjes
1Department of Urology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
Annals of Clinical Biochemistry
|March 29, 2012
Summary
Alphafetoprotein (AFP) immunoassays show method bias despite standardization. Some AFP measurement methods lack sufficient analytical performance, impacting tumor marker accuracy.
Area of Science:
- Clinical Chemistry
- Tumor Marker Analysis
- Laboratory Medicine
Background:
- Elevated alphafetoprotein (AFP) is a significant tumor marker.
- Variations in AFP assay methods can lead to differing results, even with international standardization.
- Assessing analytical performance across different AFP measurement methods is crucial.
Purpose of the Study:
- To evaluate the analytical performance of various alphafetoprotein (AFP) immunoassay methods.
- To identify method-specific biases and assess compliance with analytical performance criteria.
Main Methods:
- Analysis of annual AFP survey data from 65-75 Dutch laboratories (2005-2010).
- External Quality Assurance program data from the SKML/Binding Analysis.
- Comparison of method biases against the all-laboratory trimmed mean and assessment against the Fraser criterion.
Main Results:
- Elecsys/Modular (36%) and Immulite 2000/2500 (29%) were the most common AFP assay methods.
- Average method biases ranged up to 15% positive and 12% negative.
- Over 70% of Immulite/Elecsys users met the Fraser criterion, while over 50% of users of other methods did not.
Conclusions:
- Method bias persists in alphafetoprotein (AFP) immunoassays, despite calibration to the same international standard.
- Certain AFP assay methods demonstrate insufficient analytical performance, potentially affecting clinical interpretation.