[Evaluation of a portable device based on peripheral arterial tone in the detection of obstructive sleep apnea]
Juan Zhang1, Chun-yan Wang, Ning-yu Wang
1Department of Otorhinolaryngology Head and Neck Surgery, Beijing Chaoyang Hospital, Capital Medical University; Capital University of Medical Sciences School of Otorhinolaryngology, Beijing 100020, China.
Objective:
To assess the accuracy of a wrist-worn device (Watch-PAT 200) in the diagnosis of obstructive sleep apnea syndrome (OSAHS).
Methods:
Forty-three adult subjects with suspected OSAHS simultaneously had a standard in-laboratory polysomnogram (PSG) and wore the Watch-PAT 200 during a full-night recording. PSG sleep and respiratory events were scored according to standard criteria. The PSG recordings were blindly manually analyzed, while Watch-PAT data were scored automatically based on the algorithm developed previously.
Results:
The mean age of the subjects was (42.2 ± 12.2) years (x(-) ± s), and mean body mass index was (28.0 ± 3.9) kg/m(2). Mean PSG apnea hypopnea index (AHI) was (34.9 ± 29.9) events per hour, and mean PAT-AHI was (36.0 ± 29.2) events per hour. There was a significant correlation between PAT AHI and AHI by PSG (r = 0.931, P < 0.01). A Bland-Altman plot of PAT AHI and PSG AHI was also used to assess the accuracy of Watch-PAT 200. At lower levels of AHI, PAT tended to overestimate disease severity, while at higher levels of AHI, Watch-PAT underestimated severity. To assess sensitivity and specificity of Watch-PAT, constructed receiver operator characteristic curves using a variety of AHI threshold values (5, 15 and 30 events per hour). For AHI ≥ 5 events per hour as threshold value, the Watch-PAT diagnosing rate was 93%, and sensitivity as well as specificity were 94.7% and 80.0%. The misdiagnosis rate and missed diagnosis rate were 20.0% and 5.3%. Optimal combinations of sensitivity and specificity for the AHI threshold values (15 and 30 events per hour) were 82.6% and 100.0%, 95.0% and 95.7% respectively.
Conclusion:
The Watch-PAT 200 may offer an accurate, robust, and reliable ambulatory method for the detection of OSAHS, with minimal patient discomfort.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...