Origin and correction of magnetic field inhomogeneity at the interface in biphasic NMR samples
Bryan T Martin1, G C Chingas, Owen M McDougal
1Department of Chemistry and Biochemistry, Boise State University, Boise, ID 83725, USA. bryanmartin@u.boisestate.edu
Abstract:
The use of susceptibility matching to minimize spectral distortion of biphasic samples layered in a standard 5 mm NMR tube is described. The approach uses magic angle spinning (MAS) to first extract chemical shift differences by suppressing bulk magnetization. Then, using biphasic coaxial samples, magnetic susceptibilities are matched by titration with a paramagnetic salt. The matched phases are then layered in a standard NMR tube where they can be shimmed and examined. Linewidths of two distinct spectral lines, selected to characterize homogeneity in each phase, are simultaneously optimized. Two-dimensional distortion-free, slice-resolved spectra of an octanol/water system illustrate the method. These data are obtained using a 2D stepped-gradient pulse sequence devised for this application. Advantages of this sequence over slice-selective methods are that acquisition efficiency is increased and processing requires only conventional software.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Magnetostatic Boundary Conditions
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
Atomic Nuclei: Nuclear Relaxation Processes
Atomic Nuclei: Magnetic Resonance
Atomic Nuclei: Larmor Precession Frequency