Related Experiment Video
Updated: May 23, 2026

Active Probe Atomic Force Microscopy with Quattro-Parallel Cantilever Arrays for High-Throughput Large-Scale Sample Inspection
Published on: June 13, 2023
A novel method of temperature compensation for piezoresistive microcantilever-based sensors
Jianqiang Han1, Xiaofei Wang, Tianhong Yan
1College of Mechanical and Electrical Engineering, China Jiliang University, 258 Xueyuan Road, Xiasha High-Edu Park, Hangzhou 310018, China. hjqsmx@sina.com
Abstract:
Microcantilever with integrated piezoresistor has been applied to in situ surface stress measurement in the field of biochemical sensors. It is well known that piezoresistive cantilever-based sensors are sensitive to ambient temperature changing due to highly temperature-dependent piezoresistive effect and mismatch in thermal expansion of composite materials. This paper proposes a novel method of temperature drift compensation for microcantilever-based sensors with a piezoresistive full Wheatstone bridge integrated at the clamped ends by subtracting the amplified output voltage of the reference cantilever from the output voltage of the sensing cantilever through a simple temperature compensating circuit. Experiments show that the temperature drift of microcantilever sensors can be significantly reduced by the method.