A phantom assessment of portable imaging and radio-guided surgery systems with technetium-99m and fluorine-18
Thomas Krohn1, Frederik A Verburg, Holger Brockmann
1Department of Nuclear Medicine, University Hospital Aachen, RWTH Aachen University, Aachen, Germany. tkrohn@ukaachen.de
Aim:
To perform a detailed analysis of the performance of mobile intraoperative imaging systems and gamma probes in a phantom set-up, and compare this with a conventional gamma camera.
Methods:
Two separate experiments were performed. In the first, a modified Jaszczak phantom equipped with five (99m)Tc-filled hot spheres (0.5-20 ml) was analyzed using Sentinella, declipseSPECT and a conventional gamma camera under three conditions: no background, spheres on the surface of the background activity, and totally immersed spheres (contrast level in both 1: 8). In the second experiment, two phantom spheres (0.5 and 2 ml) filled with (99m)Tc and (18)F (infinite contrast, 1: 4 and 1: 8) were measured using the hand-held probes Navigator and GammaLocator DXI. Data analysis consisted of signal-to-background ratios and determination of the full-width at half-maximum (FWHM). A visual scoring was performed by three nuclear medicine physicians.
Results:
At infinite contrast, (99m)Tc-filled spheres with volumes of at least 2 ml could be detected adequately with all systems (e.g. 2 ml sphere, FWHM: ECAM 11 mm, declipseSPECT 9 mm, Navigator 13 mm, GammaLocator 12 mm). Under decreased contrast conditions, the results for all systems were impaired and the 0.5 ml phantom sphere filled with either (99m)Tc or (18)F was only detected accurately by the GammaLocator (FWHM range: 13-17 mm).
Conclusion:
All systems are suitable for intraoperative sentinel node detection with nearly infinite signal-to-background contrast. At a lower contrast, the GammaLocator performed best for the detection of small volumes at low-contrast ratios regardless of the radionuclide.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...