Quantitative analysis shows that contrast medium in positron emission tomography/computed tomography may cause
Clare McKeown1, Mary-Frances Dempsey, Gerry Gillen
1West of Scotland PET Centre, Gartnavel General Hospital, Glasgow, UK. clare.mckeown@nhs.net
Objectives:
Attenuation correction algorithms are required for accurate quantification of PET data and for mapping of radioactive tracers. Modern PET systems incorporate computed tomography (CT) systems to perform attenuation correction. However, high-density media, such as contrast agents, may introduce potentially clinically significant artefacts in PET images when CT-based attenuation correction algorithms are used. Although various groups have investigated this issue, no study has quantitatively assessed the clinical significance of these artefacts by comparing artefact and lesion standardized uptake values (SUVs) in controlled phantom experiments. Furthermore, previous studies have focussed on the effects of increasing the concentration of contrast medium, without investigating the effects of increasing its transaxial area. This study quantifies the clinical significance of increasing the concentration and transaxial area of contrast agents and evaluates a commercially available contrast agent correction algorithm.
Methods:
Images of a phantom containing background activity, a volume of contrast agent and varying sizes of hot lesions were acquired using clinical acquisition protocols. Quantitative analysis was performed on transaxial image slices of PET data.
Results:
The densest medium caused a 125% SUV(mean) increase in the area containing, and immediately adjacent to, contrast medium when compared with a reference water phantom. As the transaxial area of the contrast medium increased, artefacts appeared as a ring of activity around the periphery of the contrast medium. The contrast correction algorithm reduced these artefacts to within ± 39% of the reference results.
Conclusion:
Oral and IV contrast agents can cause clinically significant artefacts in CT-based attenuation-corrected PET images and should be used with caution.
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...