Variability of a closed, rebreathing setup for multiple breath wash-out testing in children
Jessica E Pittman1, Robin C Johnson, Paul W Jones
1Division of Pediatric Pulmonology, University of North Carolina at Chapel Hill, Chapel Hill, NC 27599, USA. jessica_pittman@med.unc.edu
Objective:
The multiple breath wash-out technique (MBW) that measures lung clearance index (LCI) and functional residual capacity (FRC) may be more sensitive than spirometry for identification of early obstructive airways disease. The open MBW setup using mass spectrometry referenced in previous publications is not readily available in the U.S. Our objective was to assess validity and sensitivity of a commercially available device that uses a closed (rebreathing) setup with photoacoustic spectroscopy for MBW testing.
Study Design And Methods:
Subjects aged 5-21 who were either healthy or had a history of cystic fibrosis were enrolled. Subjects completed MBW (Innocor device; Innovision, Denmark) and spirometry; measures obtained included LCI, FRC, and forced expiratory volume in 1 sec, as well as changes in end-tidal carbon dioxide levels (CO(2) ) and tidal volume during MBW testing.
Results:
Seventeen subjects attempted a total of 76 MBW maneuvers; 80% were completed and 60% met criteria for acceptability; most were unacceptable due to errors in the tracer gas curve. Substantial intra-subject variability for LCI and FRC were noted (mean 26% ± 55 and 36% ± 63, respectively). Subjects were also noted to have significant increases in exhaled CO(2) and tidal volume during MBW testing.
Conclusions:
In our initial experience using a commercially available closed setup for MBW testing, we found a significant degree of intra-subject variability leading us to suspend testing. Variability could be due to hypercapnea and instability of tidal breathing secondary to the rebreathing setup. Further studies are needed to better understand the closed system MBW setup.
Related Concept Videos
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes include...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...