Related Experiment Video
Updated: May 23, 2026

Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Quantitative assessment of shockwave lithotripsy accuracy and the effect of respiratory motion
Mathew D Sorensen1, Michael R Bailey, Anup R Shah
1Department of Urology, University of Washington School of Medicine, Seattle, Washington 98195, USA. mathews@uw.edu
Background And Purpose:
Effective stone comminution during shockwave lithotripsy (SWL) is dependent on precise three-dimensional targeting of the shockwave. Respiratory motion, imprecise targeting or shockwave alignment, and stone movement may compromise treatment efficacy. The purpose of this study was to evaluate the accuracy of shockwave targeting during SWL treatment and the effect of motion from respiration.
Patients And Methods:
Ten patients underwent SWL for the treatment of 13 renal stones. Stones were targeted fluoroscopically using a Healthtronics Lithotron (five cases) or Dornier Compact Delta II (five cases) shockwave lithotripter. Shocks were delivered at a rate of 1 to 2 Hz with ramping shockwave energy settings of 14 to 26 kV or level 1 to 5. After the low energy pretreatment and protective pause, a commercial diagnostic ultrasound (US) imaging system was used to record images of the stone during active SWL treatment. Shockwave accuracy, defined as the proportion of shockwaves that resulted in stone motion with shockwave delivery, and respiratory stone motion were determined by two independent observers who reviewed the ultrasonographic videos.
Results:
Mean age was 51 ± 15 years with 60% men, and mean stone size was 10.5 ± 3.7 mm (range 5-18 mm). A mean of 2675 ± 303 shocks was delivered. Shockwave-induced stone motion was observed with every stone. Accurate targeting of the stone occurred in 60% ± 15% of shockwaves.
Conclusions:
US imaging during SWL revealed that 40% of shockwaves miss the stone and contribute solely to tissue injury, primarily from movement with respiration. These data support the need for a device to deliver shockwaves only when the stone is in target. US imaging provides real-time assessment of stone targeting and accuracy of shockwave delivery.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: