Related Experiment Video
Updated: May 23, 2026

A Magnetic Resonance Imaging Protocol for Stroke Onset Time Estimation in Permanent Cerebral Ischemia
Published on: September 16, 2017
Rapid time-resolved magnetic resonance angiography via a multiecho radial trajectory and GraDeS reconstruction
Gregory R Lee1, Nicole Seiberlich, Jeffrey L Sunshine
1Department of Radiology, School of Medicine, Case Western Reserve University / University Hospitals of Cleveland, Cleveland, Ohio 44106, USA. gregory.r.lee@case.edu
Abstract:
Contrast-enhanced magnetic resonance angiography is challenging due to the need for both high spatial and temporal resolution. A multishot trajectory composed of pseudo-random rotations of a single multiecho radial readout was developed. The trajectory is designed to give incoherent aliasing artifacts and a relatively uniform distribution of projections over all time scales. A field map (computed from the same data set) is used to avoid signal dropout in regions of substantial field inhomogeneity. A compressed sensing reconstruction using the GraDeS algorithm was used. Whole brain angiograms were reconstructed at 1-mm isotropic resolution and a 1.1-s frame rate (corresponding to an acceleration factor > 100). The only parameter which must be chosen is the number of iterations of the GraDeS algorithm. A larger number of iterations improves the temporal behavior at cost of decreased image signal-to-noise ratio. The resulting images provide a good depiction of the cerebral vasculature and have excellent arterial/venous separation.