Related Experiment Video
Updated: May 23, 2026

Drug-Induced Sleep Endoscopy (DISE) with Target Controlled Infusion (TCI) and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Spectral entropy for assessing the depth of propofol sedation
Mi-Young Kwon1, Seung-Yun Lee, Tae-Yop Kim
1Department of Anesthesiology, National Medical Center, Seoul, Korea.
Background:
For patients in the intensive care unit (ICU) or under monitored anesthetic care (MAC), the precise monitoring of sedation depth facilitates the optimization of dosage and prevents adverse complications from underor over-sedation. For this purpose, conventional subjective sedation scales, such as the Observer's Assessment of Alertness/Sedation (OAA/S) or the Ramsay scale, have been widely utilized. Current procedures frequently disturb the patient's comfort and compromise the already well-established sedation. Therefore, reliable objective sedation scales that do not cause disturbances would be beneficial. We aimed to determine whether spectral entropy can be used as a sedation monitor as well as determine its ability to discriminate all levels of propofol-induced sedation during gradual increments of propofol dosage.
Methods:
In 25 healthy volunteers undergoing general anesthesia, the values of response entropy (RE) and state entropy (SE) corresponding to each OAA/S (5 to 1) were determined. The scores were then analyzed during each 0.5 mcg/ml- incremental increase of a propofol dose.
Results:
We observed a reduction of both RE and SE values that correlated with the OAA/S (correlation coefficient of 0.819 in RE-OAA/S and 0.753 in SE-OAA/S). The RE and SE values corresponding to awake (OAA/S score 5), light sedation (OAA/S 3-4) and deep sedation (OAA/S 1-2) displayed differences (P < 0.05).
Conclusions:
The results indicate that spectral entropy can be utilized as a reliable objective monitor to determine the depth of propofol-induced sedation.
Related Concept Videos
Stages of General Anesthesia
Parenteral Anesthetics: Overview
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Sedatives and Hypnotics Drugs: Barbiturates