High beam current operation of a PETtrace™ cyclotron for 18F- production
S Eberl1, T Eriksson, O Svedberg
1Department of PET and Nuclear Medicine, Royal Prince Alfred Hospital, Camperdown, NSW 2050, Australia. eberl@staff.usyd.edu.au
Abstract:
Upgrades and optimisation achieved 160 μA total target current operation of a GE PETtrace cyclotron in dual target mode for the routine production of [(18)F]FDG for >2 years. Approximately 900 GBq of (18)F(-) and >500 GBq of [(18)F]FDG can be produced routinely in a single production run, meeting the routine [(18)F]FDG requirements with our customer base and achieving economies of scale. Production of >1 TBq of (18)F(-) in a single run was achieved. Reliability, saturation and synthesis yields were not adversely affected.