Liquid chromatography-tandem mass spectrometric assay for the JAK2 inhibitor CYT387 in plasma
Rolf W Sparidans1, Selvi Durmus, Ning Xu
1Utrecht University, Faculty of Science, Department of Pharmaceutical Sciences, Division of Pharmacoepidemiology & Clinical Pharmacology, Universiteitsweg 99, 3584 CG Utrecht, The Netherlands. R.W.Sparidans@uu.nl
Abstract:
A quantitative bioanalytical liquid chromatography-tandem mass spectrometric (LC-MS/MS) assay for the JAK2 inhibitor CYT387 was developed and validated. Plasma samples were pre-treated using protein precipitation with acetonitrile containing cediranib as internal standard. The extract was directly injected into the chromatographic system after dilution with water. This system consisted of a sub-2 μm particle, trifunctional bonded octadecyl silica column with a gradient using 0.005% (v/v) of formic acid in a mixture of water and methanol. The eluate was transferred into the electrospray interface with positive ionization and the analyte was detected in the selected reaction monitoring mode of a triple quadrupole mass spectrometer. The assay was validated in a 0.25-1000 ng/ml calibration range. Within day precisions were 3.0-13.5%, between day precisions 5.7% and 14.5%. Accuracies were between 96% and 113% for the whole calibration range. The drug was stable under all relevant analytical conditions. Finally, the assay was successfully used to assess drug levels in mice.
Related Concept Videos
Tandem Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
MALDI-TOF Mass Spectrometry
Therapeutic Drug Monitoring: Drug Analysis Methods