Related Experiment Video
Updated: May 23, 2026

The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
[Blind tracheal intubation through two supraglottic devices: i-gel versus Fastrach intubating laryngeal mask airway
J A Sastre1, T López, J C Garzón
1Servicio de Anestesiología y Reanimación, Complejo Asistencial Universitario de Salamanca, Salamanca, España.
Background And Objectives:
The use of supraglottic devices as a means of rescue in patients difficult to intubate or ventilate has increased in the field of anaesthetics and in emergency medicine. This study is designed to evaluate the success rate of blind intubations using two supraglottic devices, the Fastrach ILMA and the i-gel mask.
Patients And Methods:
A total of 80 patients (40 per group) were included. After positioning them a leak test was performed, the glottis view was checked with a fibrobronchoscope, and an attempt was made to introduce an endotracheal tube through the device, and the procedure was repeated. Adequate ventilation was evaluated, as well as the grade of fibrobronchoscope view, the success of the intubation, and the complications observed after their use.
Results:
There were no differences in the incidence of adequate ventilation with either device. The glottis view (Brimacombe scale) was better with i-gel (77.78% versus 68.42%) at the second attempt, but not on the first. A higher percentage of intubations were achieved with the Fastrach ILMA (70% versus 40%; P=.013). The incidence of throat pain was similar with both devices, but post-operative dysphonia was more frequent with i-gel (20% versus 0; P=.0053).
Conclusions:
Both devices were equally effective in achieving adequate ventilation; however, the Fastrach ILMA enabled a higher number of intubations to be made than i-gel and with a lower incidence of post-operative dysphonia.
More Related Videos
04:30Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
Published on: August 25, 2022
04:56Laryngeal Mask Airway (LMA) Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway (SGA)
Published on: July 14, 2023
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...