Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
High intensity focused ultrasound with large aperture transducers: a MRI based focal point correction for tissue
Charles Mougenot1, Matti Tillander, Julius Koskela
1Philips Healthcare, Toronto L6C 2S3, Canada. charles.mougenot@philips.com
Purpose:
The risk of undesired tissue damage to thoracic cage, heart, and lung during MR guided HIFU ablations of breast cancer can be greatly reduced if a phased array transducer design with a lateral beam direction is used in combination with a large aperture. The disadvantage is an increased sensitivity to focus aberrations due to tissue heterogeneity. Here, the authors propose to restore the focal coherence by using a matched aperture phase correction, which is based on a noninvasively obtained tissue model.
Methods:
The method combines high resolution MRI with ultrasound wave measurements of different tissue types to determine a phase correction, which compensates focal point aberrations caused by tissue heterogeneity. 3D segmentation of tissue is used to quantify the relative proportion of each tissue type along a line running from the center of each element of the phased array to the target focal point.
Results:
For tissue types with a celerity difference of 3%, the proposed method allows to quantify the phase aberration with an accuracy of 6° ± 20° and a correlation factor R(2) = 0.95. Using the refocusing method for a complex heterogeneous phantom resulted in 95% of the maximal pressure, whereas only 70% of the maximal pressure is obtained in absence of any phase correction.
Conclusions:
Since the proposed refocusing algorithm is compatible with a standard interventional preplanning and requires only a minimal amount of processing, it presents a promising approach to compensate for aberration in heterogeneous tissues such as the human breast.