An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Detachment strength of human osteoblasts cultured on hydroxyapatite with various surface roughness. Contribution of
Petros A Kokkinos1, Petros G Koutsoukos, Despina D Deligianni
1Department of Mechanical Engineering and Aeronautics, University of Patras, Rion, Patras, Greece.
Abstract:
Hydroxyapatite (HA) has been widely used as a bone substitute in dental, maxillofacial and orthopaedic surgery and as osteoconductive bone substitute or precoating of pedicle screws and cages in spine surgery. The aim of the present study was to investigate the osteoblastic adhesion strength on HA substrata with different surface topography and biochemistry (pre-adsorption of fibronectin) after blocking of specific integrin subunits with monoclonal antibodies. Stoichiometric HA was prepared by precipitation followed by ageing and characterized by SEM, EDX, powder XRD, Raman spectroscopy, TGA, and specific surface area analysis. Human bone marrow derived osteoblasts were cultured on HA disc-shaped substrata which were sintered and polished resulting in two surface roughness grades. For attachment evaluation, cells were incubated with monoclonal antibodies and seeded for 2 h on the substrata. Cell detachment strength was determined using a rotating disc device. Cell detachment strength was surface roughness, fibronectin preadsorption and intergin subunit sensitive.
More Related Videos
09:35Distinctive Capillary Action by Micro-channels in Bone-like Templates can Enhance Recruitment of Cells for Restoration of Large Bony Defect
Published on: September 11, 2015
10:52Differentiation and Characterization of Osteoclasts from Human Induced Pluripotent Stem Cells
Published on: March 22, 2024