Assessment of Cardiac Function and Myocardial Morphology Using Small Animal Look-locker Inversion Recovery (SALLI) MRI in Rats
Published on: July 19, 2013
Multiple 3D inversion recovery imaging for volume T1 mapping of the heart
A Coniglio1, P Di Renzi, G Vilches Freixas
1AFaR, Department of Medical Physics, S. Giovanni Calibita, Fatebenefratelli Hospital, Rome, Italy. coniglio.fbf.isola@gmail.com
Abstract:
In this article, a three-dimensional inversion recovery sequence was optimized with the aim of generating in vivo volume T(1) maps of the heart using a 1.5-T MR system. Acquisitions were performed before and after gadolinium diethylenetriamine penta-acetic acid (Gd-DTPA) administration in one patient with hypertrophic cardiomyopathy and in two healthy volunteers. Data were acquired with a multishot fast field echo readout using both ECG and respiratory triggers. A dedicated phantom, composed of four solutions with different T(1) values, was positioned on the subjects' thoracic region to perform patient-specific calibration. Pixel based T(1) maps were calculated with a custom Matlab(®) code. Phantom measurements showed a good accuracy of the technique and in vivo T(1) estimation of liver, skeletal muscle, myocardium, and blood resulted in good agreement with values reported in the literature. Multiple three-dimensional inversion recovery technique is a feasible and accurate method to perform T(1) volume mapping.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies VII: Vascular Imaging