The effect of preactivation vial temperature on the acoustic properties of Definity™
Brandon L Helfield1, Xuan Huo, Ross Williams
1Department of Medical Biophysics, University of Toronto, Toronto, Ontario, Canada. brandon.helfield@sri.utoronto.ca
Abstract:
Definity™ is a widely available clinically approved ultrasound contrast agent. The manufacturer's instructions indicate that the refrigerated vial should be allowed to reach room temperature prior to its 45 s mechanical agitation activation process. Activation results in vial heating and it has been previously observed that "smaller" bubbles are formed later in this process (>10 s) when the vial temperature is elevated. The objective of this work was to examine the effects of preactivation vial temperature on the size distribution, frequency dependent attenuation (1.5-27 MHz) and nonlinear imaging performance of Definity™. Experiments were conducted with vials at refrigerator temperature (2°C), room temperature (22°C) or 37°C at the outset of the activation procedure. The size distributions were found to be strongly dependent on preactivation vial temperature and the attenuation results indicated considerable differences in the frequency response of the agent, most notably the appearance of a peak at 4 MHz for the 2°C case. Nonlinear imaging results performed using a 1-5 MHz transducer probe with a wall-less vessel phantom indicated that 2°C vials produced a signal enhancement 5.1 dB greater than for 22°C Definity™ (p < 0.05). These results, therefore, indicate that not permitting the vial to reach room temperature has a considerable impact on the imaging performance of Definity™. Conversely, activating a cooled vial can provide a means by which to improve contrast enhancement when using low frequency clinical transducers.
Related Concept Videos
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature from...
Effect of Temperature Change on Reaction Rate
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Intensity and Pressure of Sound Waves
Unlike the time average of a sinusoidal term, which is zero since it is positive and...
Temperature Dependent Deformation