Carbon dioxide rebreathing with the anaesthetic conserving device, AnaConDa®
L W Sturesson1, G Malmkvist, M Bodelsson
1Section of Anaesthesiology and Intensive Care, Lund Universityand Skane University Hospital Lund, Sweden. louise.walther_sturesson@med.lu.se
Background:
The anaesthetic conserving device (ACD) AnaConDa(®) was developed to allow the reduced use of inhaled agents by conserving exhaled agent and allowing rebreathing. Elevated has been observed in patients when using this ACD, despite tidal volume compensation for the larger apparatus dead space. The aim of the present study was to determine whether CO(2), like inhaled anaesthetics, adsorbs to the ACD during expiration and returns to a test lung during the following inspiration.
Methods:
The ACD was attached to an experimental test lung. Apparent dead space by the single-breath test for CO(2) and the amount of CO(2) adsorbed to the carbon filter of the ACD was measured with infrared spectrometry.
Results:
Apparent dead space was 230 ml larger using the ACD compared with a conventional heat and moisture exchanger (internal volumes 100 and 50 ml, respectively). Varying CO(2) flux to the test lung (85-375 ml min(-1)) did not change the measured dead space nor did varying respiratory rate (12-24 bpm). The ACD contained 3.3 times more CO(2) than the predicted amount present in its internal volume of 100 ml.
Conclusions:
Our measurements show a CO(2) reservoir effect of 180 ml in excess of the ACD internal volume. This is due to adsorption of CO(2) in the ACD during expiration and return of CO(2) during the following inspiration.
More Related Videos
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
14:52Recording Brain Electromagnetic Activity During the Administration of the Gaseous Anesthetic Agents Xenon and Nitrous Oxide in Healthy Volunteers
Published on: January 13, 2018
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Acute Respiratory Failure-III