Can the individual calibration be modified when laser fluorescence method is used for caries detection?
Alessandra Reyes1, Gisele E Ferreira, Joyce Santos
1Departmento de Ortodontia e Odontopediatria, Faculdade de Odontologia da Universidade de São Paulo, São Paulo, Brazil.
International Journal of Paediatric Dentistry
|April 20, 2012
Summary
Individual calibration (IC) for laser fluorescence (LF) caries detection is time-consuming. This study found that while no calibration increases LF readings and affects reproducibility, various IC methods yield similar results, suggesting flexibility in clinical practice.
Area of Science:
- Dentistry
- Biophotonics
- Medical Diagnostics
Background:
- Individual calibration (IC) for fluorescence-based caries detection is time-intensive, particularly for pediatric dentists when performed tooth-by-tooth.
- The precise impact of different IC modes on laser fluorescence (LF) readings for caries detection remains unclear.
Purpose of the Study:
- To investigate the influence of various individual calibration (IC) methods on laser fluorescence (LF) readings in vivo.
- To assess the impact of different calibration strategies on the accuracy and reproducibility of LF caries detection.
Main Methods:
- Ninety-six occlusal and 95 buccal surfaces of permanent molars were examined using an LF device.
- IC was performed using different modes: no calibration (control), the examined tooth, a permanent incisor, a first primary molar, or a second primary molar, all within the same child.
- Statistical analysis included Wilcoxon test, Bland-Altman analysis, and calculation of intraclass correlation coefficients (ICC).
Main Results:
- Laser fluorescence readings were significantly higher when no prior calibration was performed compared to any IC mode.
- The absence of calibration also resulted in different ICC values, indicating reduced reproducibility.
- LF readings after various IC modes (excluding no calibration) were statistically similar.
Conclusions:
- The absence of individual calibration (IC) significantly impacts laser fluorescence (LF) readings and their reproducibility in caries detection.
- Different IC methods can be considered viable options in clinical practice, offering flexibility for dentists.
- These findings suggest that while calibration is crucial, specific tooth-by-tooth calibration may not always be necessary, potentially streamlining the diagnostic process.