Presentation of a cone-beam CT scanning protocol for preprosthetic cranial bone grafting of the atrophic maxilla
J De Ceulaer1, G Swennen, J Abeloos
1Department of Surgery, Division of Maxillofacial Surgery, General Hospital St. John Bruges, Belgium. maxfac@azbrugge.be
Abstract:
The use of autologous calvarian split thickness bone grafts is a well accepted preprosthetic surgical procedure for reconstruction of the severely atrophied maxilla. Although generally accepted as being a superior bone graft regarding long-term stability, the reported risks in the literature may dissuade the surgeon. A modified cone-beam computed tomography (CBCT) scanning protocol (extended field of view starting from 1cm below the occlusal plane up to the limit of the cranial vault, 0.4 voxel) is proposed that allows assessment of both the cranial donor site as the maxillary receptor site and the sinus conditions with a single scan. Issues regarding quality of the data, radiation dose and clinical practicability are discussed.