Evaluation of an automated sphygmomanometer for use in the office setting
Martin G Myers1, Miguel Valdivieso
1Division of Cardiology, Schulich Heart Program, Sunnybrook Health Sciences Centre, Toronto, Ontario, Canada. martin.myers@sunnybrook.ca
Objective:
To determine whether the WatchBP Office sphygmomanometer can be used to obtain automated office blood pressure (AOBP) readings that are similar to the awake ambulatory BP.
Methods:
One hundred patients referred for 24 h ambulatory BP monitoring had BP recorded three times using the WatchBP Office fully automated sphygmomanometer in accordance with standard AOBP measurement guidelines. The mean AOBP was compared with the mean awake ambulatory BP.
Results:
The mean (± SD) AOBP (138.6 ± 13.7/79.7 ± 9.0 mmHg) was similar to the mean awake ambulatory BP (136.8 ± 12.4/79.0 ± 10.8 mmHg). The small difference in systolic BP (1.8 mmHg) was statistically significant (P=0.03), but was within the accepted range (5 mmHg) recommended by guidelines for equivalence between BP readings. There was a strong correlation (P<0.001) between the systolic/diastolic AOBP and awake ambulatory BP readings (r=0.819/0.801).
Conclusion:
The WatchBP Office produces BP readings that closely approximate the awake ambulatory BP, confirming that this automated sphygmomanometer is suitable for recording AOBP in clinical practice.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure