Failure mechanism of a constrained liner: a case report
Rui Niimi1, Akihiro Sudo, Masahiro Hasegawa
1Department of Orthopaedic Surgery, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu City, Mie, 514-8507, Japan. furikakefuri@hotmail.co.jp
Abstract:
Instability after primary and revision total hip arthroplasty continues to be a problem. The use of a constrained system helps manage this problem. A new constrained total hip arthroplasty, Trilogy constrained liner (Zimmer, Warsaw, IN), is currently in use. We report a case showing dislocation following a Trilogy constrained total hip arthroplasty. In this case, when an impingement between the femoral neck and the anterior part of the polyethylene liner occurred, the hip dislocated easily although both polyethylene and reinforcing ring were properly positioned. The lever-out test showed that the Trilogy constrained liner is safe and compares favourably with other implants. Surgeons should be aware that constrained acetabular systems are not infallible and they should pay attention to place implants in good position even when constrained THA is performed.
Related Concept Videos
Stress Concentrations in Circular Shafts
Eccentric Axial Loading in a Plane of Symmetry
Plastic Behavior
Constraints and Statical Determinacy
Yield Criteria for Ductile Materials under Plane Stress
The Maximum Shearing Stress Criterion, also known as the...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical bending,...