K Vörös1, A Bonnevie, J Reiczigel
1Department and Clinic Internal Medicine, Szent István University, Budapest, Hungary. Voros.Karoly@aotk.szie.hu
Background And Objective:
Cardiac auscultation is one of the most important parts of the cardiological examination traditionally performed with acoustic stethoscopes. The aim of this study was to compare the sensitivities and the diagnostic capabilities of traditional and electronic stethoscopes in detecting canine heart murmurs.
Materials And Methods:
The study was performed on 21 dogs referred for cardiologic examination with suspected heart murmurs. Six out of these dogs had cardiac murmurs bilaterally. Cardiac auscultation was performed independently by a final-year veterinary student (AB=I1) and by an experienced clinician (KV=I2), both using a traditional and a Welch Allyn Meditron electronic sensor-based stethoscope. Final diagnoses were established by echocardiography and by digital phonocardiography.
Results:
Correct detection of a murmur was made by I1 with a traditional stethoscope in 20/27 (74.0%) of the suspected murmurs (p=0.30, kappa[κ] =0.2) and with the electronic stethoscope in 26/27 (96.3%), respectively (p=0.0013, κ=0.75). I2 correctly detected the murmurs with the traditional stethoscope in 25/27 (92.6%) cases (p=0.0013, κ=0.75) and with the electronic stethoscope in all 27/27 (100%) cases (p=0.00012, κ=1). Agreements of murmur intensity gradings between traditional and electronic stethoscopes were highly significant (I1: p=6.9´10⁻⁸; κ=0.79), (I2: p=5.2´10⁻¹¹; κ=0.92). When grading the murmurs with the traditional stethoscope, there was a significant agreement between I1 and I2 (p=2.9´10⁻⁷; κ=0.79), being even higher with the electronic stethoscope (p=1.1´10⁻¹¹; κ=0.92).
Conclusion:
The electronic stethoscope was more sensitive than the traditional one in detecting and grading cardiac murmurs being especially useful for I1 with less experience. However, it can be suggested to use a traditional and an electronic stethoscopes simultaneously to optimally utilize their advantages.
Related Concept Videos
Cardiovascular System Abnormal Findings II: Auscultation
Abnormal Heart Sounds
Gallops:
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V) valves at the...
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: