Effect of beam-flow angle on velocity measurements in modern Doppler ultrasound systems
Michael Yong Park1, Seung Eun Jung, Jae Young Byun
1Department of Radiology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seocho-gu, Seoul, Korea.
Objective:
The purpose of this article is to examine the effect of different beam-flow angles on the accuracy of Doppler ultrasound velocity measurements in modern ultrasound systems.
Materials And Methods:
A flow phantom was used to create a steady flow of water in a 4.3-mm-diameter tube. Using three different modern university-grade ultrasound systems, flow was measured at 30°, 40°, 50°, 60°, 70°, 80°, and 88° beam-flow angles twice by two radiologists in consensus using a convex and linear probe. Measured flow ratio, defined as measured velocity divided by estimated actual velocity, was calculated. Intraprobe, interprobe, and intermachine mean variation of measured flow ratio were calculated.
Results:
Measured flow ratio increased as beam-flow angles increased. Measured flow ratios for the angles 30°, 40°, 50°, 60°, 70°, 80°, and 88° were 0.90, 0.97, 1.10, 1.22, 1.62, 2.34, and 10.29, respectively. Intraprobe, interprobe, and intermachine variation did not show marked differences. For angles grouped as 30-40°, 50-60°, 70°, and 80-88°, intraprobe variation was 12%, 15%, 15%, and 26%; interprobe variation was 20%, 16%, 13%, and 26%; and intermachine variation was 16%, 16%, 17%, and 54%, respectively. As beam-flow angle increased, an increase in spectral broadening was also noted.
Conclusion:
There is no simple cutoff beam-flow value, such as the well-quoted less than 60°, at which velocity measurements can be considered accurate. For follow-up imaging, beam-flow angle differences should be considered, and the same beam-flow angles should be used when possible. Follow-up imaging by different sonography machines is feasible.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Doppler Effect - I
Doppler Effect - II
Velocity and Acceleration in Steady and Unsteady Flow
The acceleration can be generalized to any point in the flow, and expressed as components along three perpendicular directions, representing changes in velocity over time.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pipe Flowrate Measurement
The orifice meter is a simple,...