Influence of different flow-triggering levels on the breathing effort of mechanically ventilated patients
1Division of Chest Medicine, Department of Internal Medicine, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan, Taiwan.
Background:
Although a sensitive flow triggering (FT) level is supposed to be associated with reduced breathing effort, the incidence of autotriggering (AT) is likely to be increased. The actual effects of various FT levels on the work of breathing and occurrence of AT in mechanically ventilated patients are unknown. We investigated the effects of different FT levels (1-8 L/min) on breathing effort and incidence of AT in mechanically ventilated patients under pressure support ventilation using a Puritan-Bennett 840 ventilator.
Methods:
Eight FT levels were randomly studied in mechanically ventilated patients under pressure support ventilation. The triggering effort (pressure-time product of triggering, PTPtr) was assessed by quantitating a segment of the pressure-time product of the esophagus (PTPes). The total PTPes, inspiratory work of breathing (Wi) and P0.1 were determined.
Results:
Nine patients with appropriately recorded signals were included. The incidence of AT significantly decreased with increasing FT level (FT1, 1 L/min: 30.7%, FT8: 0.2%). PTPtr significantly increased with increasing FT level (0.020 ± 0.004 cmH2O • S in FT1 to 0.190 ± 0.017 cmH2O • S in FT8), but P0.1 remained similar. PTPtr accounted for only1-3% of total PTPes. Wi and PTPes were significantly lower only at FT1, but there was no significant difference in Wi and PTPes at different FT levels when AT breaths were excluded.
Conclusion:
A higher FT level was associated with lower incidence of AT, but without a significant increase in breathing effort. A higher FT level may be more reasonable in mechanically ventilated patients with this particular ventilator.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: