The use and accuracy of manual and electronic gestational age calculators
Elizabeth M Smout1, Paul T Seed, Andrew H Shennan
1Department of Reproduction and Endocrinology, Maternal and Fetal Research Unit, London, UK. Elizabeth.smout@kcl.ac.uk
Objectives:
The accuracy of gestational age (GA) wheels has been shown to be poor, yet they remain commonly used. We surveyed their use within our institution and determined the inter-observer and inter-device variability of a range of devices. We have devised a procedure for validating device accuracy.
Methods:
All clinicians within our maternity unit were asked their most recent method of calculating GA and whether they felt this was accurate to within one day. Ten clinicians assessed 16 devices: 14 manual, 2 electronic. Five dates represented the last menstrual period (LMP) and were used to calculate the estimated date of delivery (EDD) of the 5 dates for each device compared to a 280 day control. ANOVA was used for statistical analysis.
Results:
73% last used a manual device to calculate GA. Seventy-two per cent believed their method was accurate. There was a significant bias (difference in the mean) between device-calculated and control EDD for all manual devices except one, with individual differences of up to 4 days. Variability altered throughout the year. Electronic devices consistently had no error. Inter-observer variability was insignificant.
Conclusions:
Clinicians should be aware of the inaccuracy of manual devices. Electronic devices are recommended. Manual devices should be validated before use by comparing the device-calculated EDD with a 280 day control at five points throughout the year.
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...