Background:
- Cochlear implants (CIs) enhance audibility for children with severe-to-profound hearing loss.
- Speech processor parameters like T-level, input dynamic range (IDR), and microphone sensitivity impact CI function.
- Optimal CI programming for pediatric audibility and speech recognition remains under-researched.
Purpose of the Study:
- To determine optimal T-level, IDR, and sensitivity settings for pediatric Advanced Bionics CI users.
- To evaluate the impact of these parameters on audibility and speech recognition in children.
Main Methods:
- Two experiments were conducted with 11 children (ages 7-17) using Advanced Bionics CIs.
- Experiment 1 compared T-levels set at 10% M-levels versus 10 CUs below "soft" for sound detection.
- Experiment 2 assessed IDR/sensitivity combinations (50/0, 50/10, 70/0, 70/10) for speech recognition in quiet and noise.
Main Results:
- Setting T-levels 10 CUs below "soft" significantly improved detection of Ling sounds and FM tones.
- A 50 dB IDR with 0 sensitivity showed poorer thresholds for low-frequency sounds and voiced Ling sounds.
- No significant group differences in speech recognition (CNC words, HINT-C) were found across IDR/sensitivity settings, but individual results varied in noise.
Conclusions:
- Manually setting T-levels based on loudness judgments is recommended for pediatric CI users, especially for inaudible low-level sounds.
- Individualized programming and verification of speech recognition in both quiet and noise are crucial.