Related Experiment Video
Updated: May 22, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Comparison of cross-sections of different femoral components for revision total knee replacement
Carmen Zietz1, Philipp Bergschmidt, Andreas Fritsche
1Department of Orthopaedics, University of Rostock, Rostock, Germany. carmen.zietz@med.uni-rostock.de
Unlabelled:
PURPOSE. To compare the inner contour of the femoral component of 10 total knee replacement (TKR) designs for possible exchange in use.
Methods:
Inner contours of the femoral components of 10 cemented, cruciate-retaining TKR designs (e.motion, Genesis, Genia, Innex, LCS, Multigen Plus, NexGen, PFC, Scorpio, Vanguard) were scanned and reconstructed to 2-dimensional contours. Their cross-sections were compared by superimposition and aligning at the distal and anterior cuts. The patellar notch and outer contour were not analysed.
Results:
The maximum deviation was 5 mm in the posterior and posterior oblique cuts and 10 mm in the anterior oblique cut. Based on similarity of the inner contour, LCS and Innex was classified as group I, e.motion, Genesis, Scorpio, Vanguard, and Multigen Plus as group II, and Genia, NexGen, and PFC as group III. All 2 designs in group I were not compatible with the other 8 designs. Four of the 5 designs in group II showed good compatibility. All 3 designs in group III significantly differed in the posterior and oblique cuts.
Conclusion:
A standardised inner contour of the femoral component can increase compatibility of different TKR systems in revision surgery and reduces the extent of bone resection.
Related Concept Videos
Bones of the Lower Limb: Femur and Patella
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris group...