Published on: December 4, 2011
Evolution of visual performance in 70 eyes implanted with the Tecnis(®) ZMB00 multifocal intraocular lens
Carlos Palomino Bautista1, David Carmona González, Alfredo Castillo Gómez
1Hospital Universitario Quirón, Madrid, Spain.
Purpose:
To evaluate the development of visual acuity, dysphotopsia phenomena, and subjective sensations in patients implanted with the Tecnis(®) ZMB00 (Advanced Medical Optics Inc, Santa Ana, CA) multifocal intraocular lens.
Methods:
In a sample of 70 eyes, distance and near visual acuity, with and without correction, contrast sensitivity, and patient satisfaction, were analyzed at 15, 30, and 60 days post-surgery.
Results:
Mean uncorrected distance visual acuity (logarithm of minimum angle of resolution) at 15, 30, and 60 days was 0.194 ± 0.054, 0.119 ± 0.026, and 0.076 ± 0.014 (P < 0.0001), respectively, and with correction, 0.051 ± 0.007 vs 0.041 ± 0.004 vs 0.022 ± 0.002 (P < 0.0001), respectively. At 60 days, 94.3% of eyes could read 1.00 close-up without correction. Patient satisfaction in terms of dysphotopsia effects and visual acuity was excellent. The mean contrast sensitivity was 1.64 ± 0.10 (logarithmic units) measured with the Pelli-Robson test.
Conclusion:
This type of multifocal intraocular lens was very effective at the distances analyzed, producing excellent objective and subjective results.